Section 7: SCPI command reference Series 2280 Precision DC Power Supply Reference Manual
7-36 077085503 / March 2019
:CALibration:PROTected:CODE
This command changes the calibration password.
Type
Affected by
Where saved
Default value
Command only
Recall settings
Nonvolatile memory
'KI002280'
Usage
:CALibration:PROTected:CODE <string>
<string>
Calibration password: An 8-character string including letters and numbers.
Details
Before changing the password, you must first enable calibration mode by using the following
command:
:CAL:PROT:STAT ON,<password>
Example
:CAL:PROT:CODE '12345678'
Change the calibration password to "12345678".
Also see
:CALibration:PROTected:STATe (on page 7-44)
:CALibration:PROTected:COUNt?
This query command returns the number of times that the instrument has completed calibration.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Nonvolatile memory
Not applicable
Usage
:CALibration:PROTected:COUNt?
Details
Returns the number of times that the instrument has completed calibration. The calibration count will
increase after saving the calibration constants.
Example
:CAL:PROT:COUN?
Return the number of times the calibration was
completed.
Also see
None