Series 2280 Precision DC Power Supply Reference Manual Section 7: SCPI command reference
077085503 / March 2019 7-59
:OUTPut[:STATe]
This command enables or disables the source output.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
0 (OFF)
Usage
:OUTPut[:STATe] <state>
:OUTPut[:STATe] <state>,<channel>
:OUTPut[:STATe]?
<state>
Turn source off: OFF or 0
Turn source on: ON or 1
Disable source: DISable or 2
<channel>
The channel number (default value is CH1):
ï‚§ Channel 1: CH1
Channel 2: CH2
Details
Measurements cannot be made while the source is off.
Example
:OUTP ON
Switch the source output of the instrument to ON.
Also see
None