Series 2280 Precision DC Power Supply Reference Manual Section 7: SCPI command reference
077085503 / March 2019 7-19
ARM subsystem
The commands in this subsystem configure and control the trigger model for the arm layer.
:ARM[:SEQuence[n]]:COUNt
This command sets the number of times to repeat the Arm event.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
1
Usage
:ARM[:SEQuence[n]]:COUNt <count>
:ARM[:SEQuence[n]]:COUNt?
:ARM[:SEQuence[n]]:COUNt? DEFault
:ARM[:SEQuence[n]]:COUNt? MINimum
:ARM[:SEQuence[n]]:COUNt? MAXimum
[n]
Channel number: For single-channel instruments, the value is 1
<count>
The number of times to repeat the arm event: 1 to 2500
Example
ARM:COUN 4
Set the arm count to four.
Also see
:TRIGger[:SEQuence[n]]:COUNt (on page 7-149)
Trigger model (on page 3-25)