[:SOURce[n]]:LIST:END:ZERO
This command specifies the source values when the instrument is stopping a list.
Recall settings
Instrument reset
Power cycle
Usage
[:SOURce[n]]:LIST:END:ZERO <value>
[:SOURce[n]]:LIST:END:ZERO?
Channel number: For single-channel instruments, the value is 1.
When the list is stopped,
ï‚§ Use the last source values: OFF or 0
ï‚§ Set the voltage to 0 V, current source to 0.1 A: ON or 1
Details
This command allows you to end a specified list with the last source value or zero value.
Example
List ends with the 0 V and 0.1000 A.
Also see
None
[:SOURce[n]]:LIST:HTIMe
This command determines the hold time of the list running.
Usage
[:SOURce[n]]:LIST:HTIMe <type>
[:SOURce[n]]:LIST:HTIMe?
Channel number: For single-channel instruments, the value is 1
POINt: The list runs until the reading reaches the preset sample count; the hold time
is computed by sample count * sample speed; the preset dwell time is ignored in
this mode
TIME: The list runs until the preset time is reached; the preset sample count is
ignored in this mode
Hold time is decided by whichever is longer: dwell time or sample count.