Section 7: SCPI command reference Series 2280 Precision DC Power Supply Reference Manual
7-74 077085503 / March 2019
SOURce subsystem
The commands in the SOURce subsystem configure and control the current source and voltage
source.
[:SOURce[n]]:DELay
This command sets the source delay.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
0.001 s
Usage
[:SOURce[n]]:DELay <value>
[:SOURce[n]]:DELay?
[n]
Channel number: For single-channel instruments, the value is 1
<value>
Delay: 0 to 99 s. Resolution 0.00001 s.
Details
This command sets a delay for the source. After the programmed source is turned on, this delay
allows the source level to settle before a measurement is taken. For details, refer to Source delay (on
page 3-7).
This delay is the same for both the current and voltage source. Delay time must be either zero or a
positive value. If you attempt to set it to a negative value, an error -222, "Data out of range" is
generated.
Example
:DELay 0.05
Set the delay for the output source to 50 ms.
Also see
None

Table of Contents

Other manuals for Keithley 2280 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2280 Series and is the answer not in the manual?

Keithley 2280 Series Specifications

General IconGeneral
BrandKeithley
Model2280 Series
CategoryPower Supply
LanguageEnglish

Related product manuals