Section 7: SCPI command reference Series 2280 Precision DC Power Supply Reference Manual
7-124 077085503 / March 2019
:SYSTem:POSetup:STATe
This command enables or disables default setup loading. When it is set to On, the power setup loads setup 1 right
after it is powered on.
Usage
:SYSTem:POSetup:STATe <state>
:SYSTem:POSetup:STATe?
Disable default setup loading: OFF or 0
Enable default setup loading: ON or 1
Example
:SYST:POS:STAT ON
:SYST:POS:STAT?
The default setup is configured to load
automatically when the instrument is turned on.
Also see
*SAV (on page 7-7)
:SYSTem:PRESet
This command returns the instrument to the factory default conditions and erases all user data stored in
nonvolatile memory.
Usage
:SYSTem:PRESet
Details
Sending this command will return the instrument to the factory default conditions and erases all user
data stored in nonvolatile memory, such as list parameter and setup information.
This command does not influence the status structure registers and error queue.
This command resets GPIB address and IP address to default settings. After sending this command,
make sure to use the new address to communicate with the instrument.
You should allow ample time for the command to be processed before attempting to communicate
with the instrument again.