Series 2280 Precision DC Power Supply Reference Manual Section 7: SCPI command reference
077085503 / March 2019 7-7
*RST
This command resets the instrument settings to their default values.
Usage
*RST
Details
Returns the instrument to default settings, cancels all pending commands, and cancels the response
to any previously received *OPC and *OPC? commands.
Also see
Reset default values (on page 2-88)
*SAV
This command saves the present instrument settings as a user-saved setup.
Usage
*SAV <n>
Details
Save the present instrument settings as a user-saved setup. You can restore the settings with the
*RCL command.
Any command that is affected by the*RST command can be saved with the *SAV command.
Any settings that were stored previously as <n> are overwritten.
If you view scripts from the front panel of the instrument, these are stored as SETUP<n>.
Example
Saves the instrument settings in memory location 2.
Also see
*RCL (on page 7-6)