EasyManuals Logo

Keithley 2280 Series Reference Manual

Keithley 2280 Series
457 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #303 background imageLoading...
Page #303 background image
Section 7: SCPI command reference Series 2280 Precision DC Power Supply Reference Manual
7-102 077085503 / March 2019
:STATus:OPERation:INSTrument:ENABle
This command sets or reads the contents of the Operation Instrument Enable Register of the status model.
Type
Affected by
Where saved
Default value
Command and query
Power cycle
Not applicable
0
Usage
:STATus:OPERation:INSTrument:ENABle <n>
:STATus:OPERation:INSTrument:ENABle?
[n]
The status of the Operation Instrument Enable Register
Details
This command sets or reads the contents of the Enable register of the Operation Instrument Event
Register.
When one of these bits is set and the corresponding bit in the Operation Instrument Event Register is
set, the ISUM bit in the Operation Event Register is set.
When sending binary values, preface <n> with #b. When sending hexadecimal values, preface <n>
with #h. No preface is needed when sending decimal values.
Example
:STAT:OPER:INST:ENAB #b0000000000000010
Sets bit 1 of the Operation Instrument
Enable Register using a binary value.
You could also send the decimal value 2
or the hexadecimal value #h2.
Also see
None

Table of Contents

Other manuals for Keithley 2280 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2280 Series and is the answer not in the manual?

Keithley 2280 Series Specifications

General IconGeneral
BrandKeithley
Model2280 Series
CategoryPower Supply
LanguageEnglish

Related product manuals