Section 7: SCPI command reference Series 2280 Precision DC Power Supply Reference Manual
7-60 077085503 / March 2019
SENSe subsystem
The SENSe subsystem commands configure and control the measurement functions of the Series
2280.
Many of these commands are set for a specific function (concurrent, current, or voltage). For
example, you can program a range setting for each function. The settings are saved with that
function.
:SENSe[n]:FUNCtion
This command selects which type of measurement is active: Concurrent, current, or voltage.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
"CONC"
Usage
:SENSe[n]:FUNCtion <name>
:SENSe[n]:FUNCtion?
[n]
Channel number: For single-channel instruments, the value is 1
<name>
A string that contains the measurement function to make active:
ï‚§ Concurrent: "CONCurrent[:DC]"
ï‚§ Current: "CURRent[:DC]"
ï‚§ Voltage: "VOLTage[:DC]"
For details, see Select a measurement function (on page 2-74).
Details
Set this command to the type of measurement you want to make.
Reading this command returns the measure function that is presently active.
Examples
:SENS:FUNC "VOLT"
Make the voltage measurement function the active
function.
Also see
None

Table of Contents

Other manuals for Keithley 2280 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2280 Series and is the answer not in the manual?

Keithley 2280 Series Specifications

General IconGeneral
BrandKeithley
Model2280 Series
CategoryPower Supply
LanguageEnglish

Related product manuals