Section 3: Functions and features Series 2280 Precision DC Power Supply Reference Manual
3-44 077085503 / March 2019
List running time
Depending on the hold time you select in the front-panel list options, there are two types of device
actions during list running, shown in the following figures.
Figure 73: Device action in time mode
For list-running time in Time mode:
Measurement time = Dwell - Source Delay
Sample Count = Measurement Time / Sample Speed