Series 2280 Precision DC Power Supply Reference Manual Section 7: SCPI command reference
077085503 / March 2019 7-129
The options for <bufferElements> are described in the following table.
The measurement reading based on the
measurement function.
The unit of measure reading, source value, and
timestamp.
The date for the data point.
The time for the data point.
The timestamp for the data point.
The relative time when the data point was measured,
formatted in seconds.
The relative time when the data point was measured.
The reading number indicates the point index in the
trace buffer.
The RELative and RSTamp elements are defined as the time since the first reading was stored in
the buffer. These values are reset every 24-hours. The DATE, TIME, and TSTamp elements
represent absolute time from the internal real-time clock (RTC).
Example
:DATA:DATA? "READ,SOUR,REL"
Return the measurement value, source value, and
the relative time.
Output:
+1.000000E+00,+1.779709E-03,
+0.000000E+00,+1.000000E+00,
+1.735228E-03,+6.100000E-02