Model 2651A High Power System SourceMeter® Instrument Reference Manual Section 8: Troubleshooting guide
2651A-901-01 Rev. A / March 2011 8-3
Error summary
Error number Error level Error Message
-121 RECOVERABLE Invalid Character In Number
-120 RECOVERABLE Numeric Data
-109 RECOVERABLE Missing Parameter
-108 RECOVERABLE Parameter Not Allowed
-105 RECOVERABLE Trigger Not Allowed
-104 RECOVERABLE Data Type
-101 RECOVERABLE Invalid Character
0 NO_SEVERITY Queue Is Empty
603 RECOVERABLE Power On State Lost
702 FATAL Unresponsive digital FPGA
802 RECOVERABLE Output Blocked By Interlock
820 RECOVERABLE Parsing Value
900 FATAL Internal System
1100 RECOVERABLE Command Unavailable
1101 RECOVERABLE Parameter Too Big
1102 RECOVERABLE Parameter Too Small
1103 RECOVERABLE Max Greater Than Min
1104 RECOVERABLE Too many digits for param type
1106 RECOVERABLE Battery Not Present
1107 RECOVERABLE Cannot modify factory menu
1108 RECOVERABLE Menu name does not exist
1109 RECOVERABLE Menu name already exists
1110 FATAL Catastrophic analog supply failure
1200 RECOVERABLE TSPlink initialization failed
1201 RECOVERABLE TSPlink initialization failed
1202 RECOVERABLE TSPlink initialization failed
1203 RECOVERABLE TSPlink initialization failed (possible loop in node chain)
1204 RECOVERABLE TSPlink initialization failed
1205 RECOVERABLE TSPlink initialization failed (no remote nodes found)
1206 RECOVERABLE TSPlink initialization failed
1207 RECOVERABLE TSPlink initialization failed
1208 RECOVERABLE TSPlink initialization failed
1209 RECOVERABLE TSPlink initialization failed
1210 RECOVERABLE TSPlink initialization failed (node ID conflict)
1211 RECOVERABLE Node %u is inaccessible
1212 RECOVERABLE Invalid node ID
1400 RECOVERABLE Expected at least %d parameters
1401 RECOVERABLE Parameter %d is invalid
1402 RECOVERABLE User scripts lost
1403 RECOVERABLE Factory scripts lost
1404 RECOVERABLE Invalid byte order
1405 RECOVERABLE Invalid ASCII precision
1406
RECOVERABLE Invalid data format
1500 RECOVERABLE Invalid baud rate setting
1501 RECOVERABLE Invalid parity setting
1502 RECOVERABLE Invalid terminator setting
1503
RECOVERABLE Invalid bits setting

Table of Contents