EasyManuals Logo

Keithley 2651A Reference Manual

Keithley 2651A
800 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #729 background image
Model 2651A High Power System SourceMeter® Instrument Reference Manual Appendix B: Calibration
2651A-901-01 Rev. A / March 2011 B-29
16. Turn off the output:
smua.source.output = smua.OUTPUT_OFF
17. Set the source output to the negative full-scale value, for example:
smua.source.leveli = -90e-3
18. Turn on the output:
smua.source.output = smua.OUTPUT_ON
19. Allow the readings to settle, then get both the multimeter and Model 2651A current readings
at the negative full-scale output value (the Model 2651A measurement is not necessary if this
calibration step is being done on the CALA sense mode or 100 A range). The two
measurements should be made as close as possible in time. Use this command for the Model
2651A:
FS_rdg = smua.measure.i()
20. Turn off the output:
smua.source.output = smua.OUTPUT_OFF
21. Send the source calibration command using the -range, -zero and -FS multimeter readings,
and -zero and -FS source values for the parameters:
smua.source.calibratei(-range, src_Z, DMM_Z_rdg, src_FS, DMM_FS_rdg)
Where:
-range
= The negative of the present calibration range
src_Z
= The zero Model 2651A source output value
DMM_Z_rdg
= The zero DMM measurement
src_FS
= The FS Model 2651A source output value
DMM_FS_rdg
= The FS DMM measurement
Typical values for the 100 mA range:
smua.source.calibratei(-100e-3, -1e-30, -1e-6, -90e-3, -89.2e-3)
22. If this step is not on the CALA sense mode or 100 A range, send the measure calibration
command using the multimeter and Model 2651A readings, and range setting for the
parameters:
smua.measure.calibratei(-range, Z_rdg, DMM_Z_rdg, FS_rdg, DMM_FS_rdg)
Where:
-range
= The negative of the present calibration range
Z_rdg
= The zero Model 2651A measurement
DMM_Z_rdg
= The zero DMM measurement
FS_rdg
= The FS Model 2651A measurement
DMM_FS_rdg
= The FS DMM measurement
Typical 100 mA range values:
smua.measure.calibratei(-100e-3, -1e-5, -1e-6, -91e-3, -89.2e-3)

Table of Contents

Other manuals for Keithley 2651A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2651A and is the answer not in the manual?

Keithley 2651A Specifications

General IconGeneral
Maximum Voltage40 V
Current Range0 to 3 A
Resolution6.5 digits
Resolution (Current)100 pA
Minimum Current Resolution100 pA
Power300 W
Display5.7-inch color LCD
TypeSourceMeter
Communication InterfacesUSB, GPIB, Ethernet
InterfacesGPIB, USB, Ethernet
Accuracy (Voltage)0.02% + 3 mV
Resolution (Voltage)100 nV
Minimum Voltage Resolution100 nV

Related product manuals