Model 2651A High Power System SourceMeter® Instrument Reference Manual Appendix B: Calibration
2651A-901-01 Rev. A / March 2011 B-33
If you do not wish to set a calibration date or calibration due date and want to clear the previous
values, use the following commands:
smua.cal.date = 0
smua.cal.due = 0
The actual year, month, and day, as well as (optional) hour, and minute should be used (seconds can
be given but will essentially be ignored due to the precision of the internal date storage format). The
allowable range for the year is from 2005 to 2037, the month is from 1 to 12, and the day is from 1 to
31.
Step 7. Save calibration constants
Calibration is now complete, so you can store the calibration constants in nonvolatile memory by
sending the following command:
smua.cal.save()
NOTE
Unless you send the save command, the calibration you just performed will be temporary.
Step 8. Lock out calibration
To lock out further calibration, send the following command after completing the calibration
procedure:
smua.cal.lock()

Table of Contents

Other manuals for Keithley 2651A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2651A and is the answer not in the manual?

Keithley 2651A Specifications

General IconGeneral
BrandKeithley
Model2651A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals