Keysight E4980A/AL User’s Guide -437
Error Messages
Error Messages
-
-222 Data out of range
A data element (not violating the standard) outside the range defined by this
instrument has been received. This error occurs, for example, when an
integer-based command for which the parameter can be rounded exceeds the
range of -2147483648 to +2147483647 or when a real-number-based
command for which the parameter can be rounded exceeds the range of
-9.9e37 to +9.9e37. This error also occurs when a numeric value other than a
specified one is entered into a command in which the “BIN number” and “list
table number” are specified as parameters and thus the parameters are not
rounded.
-220 Parameter error
When a parameter-related error other than Errors -221 through -229 occur,
this error message is displayed.
-211 Trigger ignored
The instrument receives and detects a trigger command (“:TRIG”) or an
external trigger signal, but it is ignored due to the timing condition (the
instrument is not in the wait-for- trigger state, for example).
Suggestion:
Change the setup so that a trigger command or an external trigger signal can
be sent after the instrument has entered the wait-for- trigger state.
-200 Execution error
An error associated with execution has been generated for which this
instrument cannot specify the error message. This code indicates that an error
associated with execution defined in 11.5.1.1.5, IEEE488.2 has occurred.
-178 Expression data not allowed
Expression data is not allowed.
-171 Invalid expression
Invalid expression was received (for example, illegal character in expression).
-170 Expression error
Expression is improper.
-168 Block data not allowed
Block data is not allowed.
-161 Invalid block data
Invalid block data was received (for example, END received before length
satisfied).
-160 Block data error
Block data is improper.