Keysight E4980A/AL User’s Guide -441
Error Messages
Error Messages
-
23 Fixture OPEN det. defective
The opening and closing detection sensor of the 42842A/B protective cover
detects an error.
Suggestion:
When this error occurs, contact Keysight Technology’s Sales and Service Office
or the company from which you bought the device.
24 Fixture cover open
The protective cover of the 42842A/B is open.
Suggestion:
Close the cover.
25 DC bias I source overload
DC bias source current is overloaded.
Suggestion:
When this error occurs, lower the value of the current setting.
26 DC bias I sink overload
DC bias sink current is overloaded.
Suggestion:
When this error occurs, lower the value of the current setting.
27 DC bias unit disconnected
The 42841A is not connected.
Suggestion:
Confirm the connection to the 42841A.
28 DC bias unit connection changed
The connection with the 42841A has been changed.
Suggestion:
Cycle the power of the 42841A, then preset the E4980A/AL.
29 Cycle power of 42841A and Preset
The connection was not established with the 42841A.
Suggestion:
Cycle the power of the 42841A, then preset the E4980A/AL.