-442 Keysight E4980A/AL User’s Guide
Error Messages
Error Messages
-
30 Function type not available
The Ls-Rdc, Lp-Rdc and Vdc-Idc are not available when the 42841A is
connected.
Suggestion:
Confirm 42841A is disconnected.
35 I BIAS I/F disabled
This error occurs when DC bias current is selected when the DC bias interface
board (option 002) is not available in a model that does not have the Power/DC
bias enhance option (option 001).
Suggestion:
Press System button in the front panel and check the OPTION.
41 Correction measurement aborted
Correction data measurement is aborted.
43 Measurement failed
A measurement error occurred during correction data measurement. The OPEN
or SHORT state connected to the UNKNOWN terminals may be inappropriate.
Suggestion:
See
“Clearly abnormal measured value” on page 428.
45 Valid in single mode only
SINGLE channel correction data is measured when the MULTI channel
correction mode is set.
51 Inconsistent limit setting
For the comparator function, user attempts to set the threshold in the
sequential mode by sending the COMP:SEQ:BIN command, although the
tolerance mode is now selected.
Suggestion:
Select the sequential mode using COMP:MODE.
60 No values in sweep list
List sweep measurement is performed when the measurement point is not set
on the LIST SWEEP SETUP screen.
Suggestion:
Set the measurement point on the LIST SWEEP SETUP screen.