72 Keysight E4980A/AL Precision LCR Meter
Display Format
MEAS DISPLAY Page
3-
Conditions where overload occurs for DCR measurement
Conditions differ depending on the DCR impedance-range setting (DCR RNG
field).
3 k 2700 or more less than 2700
10 k 9,000 or more less than 9,000
30 k 27,000 or more less than 27,000
100 k 90,000 or more less than 90,000
Table 3-23 Overload range when the impedance range is in the HOLD state for
impedance measurement
Impedance range
( ) (RANGE field)
Measurable range Range in which overload occurs
Impedance
range
Conditions
AUTO When the E4980A/AL’s internal detector detects overload
HOLD Conditions differ depending on the impedance range. Refer to Table
3-24
Table 3-24 Overload range when the impedance range is in the HOLD state for DCR
measurement
Impedance range ( )
(DCR RNG field)
Measurable range Range in which overload
occurs
10 0 to 11 more than 11
100 0 or more n.a.
1 k 900 or more less than 900
10 k 9,000 or more less than 9,000
100 k 90,000 or more less than 90,000