EasyManua.ls Logo

Keysight InfiniiVision HD3 Series - Page 138

Keysight InfiniiVision HD3 Series
368 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
138 Keysight InfiniiVision HD3-Series Oscilloscopes User's Guide
10 Triggers
3 Select the trigger Source:
Analog channel, 1 to the number of channels
External — triggers on the AUX TRIG IN signal.
Line — triggers at the 50% level of the rising or falling edge of the AC power
source signal.
WaveGen — triggers at the 50% level of the rising edge of the waveform
generator output signal. (Not available when the DC, Noise, or Cardiac
waveforms are selected.)
WaveGen Mod (FM) — when waveform generator FM modulation is used,
triggers at the 50% level of the rising edge of the modulating signal.
Digital channel (on mixed-signal oscilloscopes), D0 to the number of digital
channels minus one.
You can choose a channel that is turned off (not displayed) as the source for the
edge trigger.
The selected trigger source is indicated in the upper-right corner of the display
next to the slope symbol:
1 through 4 = analog channels.
D0 through Dn = digital channels.
E = External trigger input.
L = Line trigger.
W = Waveform generator.
4 In the Slope drop-down list, select:
Rising edge.
Falling edge.
Alternating edges — Useful when you want to trigger on both edges of a
clock (for example, DDR signals).
Either edge — Useful when you want to trigger on any activity of a selected
source.
Rising and Falling edge modes operate up to the bandwidth of the
oscilloscope. Other modes operate up to the bandwidth of the oscilloscope or
1 GHz, whichever is smaller.
The selected slope is displayed in the trigger badge.

Table of Contents

Related product manuals