56 Keysight InfiniiVision HD3-Series Oscilloscopes User's Guide
3 Vertical Controls
The ground level of the signal for each displayed analog channel is identified by
the position of the icon at the far-left side of the display.
To turn waveforms on or off
Press an analog channel key turn the channel on or off.
When a channel is on, its key is lit.
On the touchscreen, you can quickly turn waveforms on or off using the
Quick Selection button and dialog box. Also, dialog boxes under the Sources >
menu have controls for turning waveforms on or off.
Channel scale,
offset
Channel 1
ground
level
Trigger
source
Trigger level
or threshold
Channel 2
ground
level
Trigger
level
Waveform quick
selection (on/off)
NOTE
Turning channels off
A channel must be selected for vertical knob adjustment—as shown by the LED next to the
channel key—before pressing the channel key will turn off the channel. For example, if channel
1 and channel 2 are turned on and the vertical knobs are enabled for channel 2, to turn
channel 1 off, press [1] to enable the vertical knobs for channel 1; then, press [1] again to turn
channel 1 off.