Save/Email/Recall (Setups, Screens, Data) 21
Keysight InfiniiVision HD3-Series Oscilloscopes User's Guide 271
To recall arbitrary waveforms
Arbitrary waveform files can be recalled from internal (\User Files) locations or
from an external USB storage device.
When recalling arbitrary waveforms (from an external USB storage device) that
were not saved from the oscilloscope, be aware that:
If the file contains two columns, the second column is automatically chosen.
If the file contains more than two columns, you are prompted to select which
column to load. Up to five columns are parsed by the oscilloscope; columns
above the fifth are ignored.
The oscilloscope uses a maximum of 8192 points for an arbitrary waveform. For
more efficient recalls, make sure your arbitrary waveforms are 8192 points or
less.
To recall an arbitrary waveform:
1 From the main menu, choose File > Open....
2 In the Open dialog box, from the Format drop-down menu, select Arbitrary
Waveform data (*.csv).
3 Use the controls in the bottom of the dialog box to select the file to be recalled.
4 Select Open.
See Also
"To edit arbitrary waveforms" on page 254
Recalling Default Setups
1 From the main menu, choose Control > Default....
2 In the Default dialog box, you can select:
Undo Default Setup — Returns to the oscilloscope settings that were present
before the last Default Setup.
Default Setup — Restores the oscilloscope's default settings. This places the
oscilloscope in a known operating condition. See
"Recall the Default
Oscilloscope Setup"
on page 27.
This is the same as pressing the front panel [Default Setup] key.
Some user settings are not changed when recalling the default setup.
Default GUI Layout — Returns to the default graphical user interface (GUI)
layout.
Factory Default — Recalls the oscilloscope's factory default settings.
You must confirm the recall because there are no user settings that are left
unchanged.

Table of Contents