202 Keysight InfiniiVision HD3-Series Oscilloscopes User's Guide
14 Measurements
1 From the menu, choose Measure > Measurements....
2 In the Measurements dialog box, select the Add, Edit & Delete tab.
3 From the Type list, select the Phase measurement.
4 In the Add Parameter area, select:
Source 1, Source 2 — Waveform sources.
5 Select Add Phase to activate the measurement.
The example below shows a phase measurement between the channel 1 and the
math d/dt function on channel 1.
X at Max Y
X at Max Y is the X axis value (usually time) at the first displayed occurrence of the
waveform Maximum, starting from the left-side of the display. For periodic signals,
the position of the maximum may vary throughout the waveform. The X marker
shows where the current X at Max Y value is being measured.
Phase =
Delay
Source 1 Period
× 360
Source 2
Source 1
Delay
Period