23 Web Interface
Accessing the Web Interface / 294
Control Instrument / 295
Get Image / 297
Save/Recall / 298
Identification Function / 300
Instrument Utilities / 301
Setting a Password / 302
When Keysight InfiniiVision HD3-Series oscilloscopes are set up on the LAN, you
can access the oscilloscope's built-in web server using a web browser. The
oscilloscope's web interface lets you:
• View information about the oscilloscope like its model number, serial number,
host name, IP address, and VISA (address) connect string.
• Send SCPI (Standard Commands for Programmable Instrumentation) remote
programming commands via the SCPI Commands applet window.
• Control the oscilloscope's graphical user interface using a remote VNC (Virtual
Network Computing) client.
• Get screen images and save or print them from the browser.
• Save setup files, mask files, screen images, or Reference waveform data.
• Recall setup files, mask files, or Reference waveform data.
• Activate the Identification function to identify a particular instrument by
causing a message to be displayed or a front panel light to blink.
• View installed options, view firmware versions and install firmware upgrade
files, install license files, and view calibration status (via the Instrument Utilities
page).
• View and modify the oscilloscope's network configuration.
The web interface for InfiniiVision HD3-Series oscilloscopes also provides help for
each of its pages.
Before you can use the web interface, you must place the oscilloscope on the
network and set up its LAN connection.