28 Keysight InfiniiVision HD3-Series Oscilloscopes User's Guide
1 Getting Started
Use Autoscale
Use [Auto Scale] to automatically configure the oscilloscope to best display the
input signals.
1 Press [Auto Scale].
You should see a waveform on the oscilloscope's display similar to this:
2 If you want to return to the oscilloscope settings that existed before, select the
Undo button.
3 If you want to enable "fast debug" autoscaling, change the channels
autoscaled, or preserve the acquisition mode during autoscale, use the
Autoscale dialog box (from the menu, choose Utilities > User Options..., then
select Autocale...). See
"To set Autoscale preferences" on page 280.
If you see the waveform, but the square wave is not shaped correctly as shown
above, perform the procedure
"Compensate Passive Probes" on page 29.
If you do not see the waveform, make sure the probe is connected securely to the
front panel channel input BNC and to the Probe Comp terminal.
How Autoscale Works
Autoscale analyzes any waveforms present at each channel and at the external
trigger input. This includes the digital channels, if connected.