50 Keysight InfiniiVision HD3-Series Oscilloscopes User's Guide
2 Horizontal Controls
For information about using zoom mode for measurements, refer to "To isolate a
pulse for Top measurement"
on page 192 and "To isolate an event for frequency
measurement" on page 197.
To change the horizontal scale coarse/fine adjustment setting
You can do one of these things:
Push the horizontal scale knob to toggle between fine and coarse adjustment
of the horizontal scale.
Select the Fine option in the Horizontal dialog box (that you can open by
selecting the Horizontal badge or by choosing Setup > Horizontal... from the main
menu).
When Fine is enabled, turning the horizontal scale knob changes the time/div
(displayed in the status line at the top of the display) in smaller increments. The
time/div remains fully calibrated when Fine is on.
When Fine is turned off, the Horizontal scale knob changes the time/div setting in
a 1-2-5 step sequence.
To position the time reference (left, center, right, custom)
Time reference is the reference point on the display for delay time (horizontal
position).
1 Open the Horizontal dial box by selecting the Horizontal badge or by choosing
Setup > Horizontal... from the main menu.
2 In the Horizontal Menu, press Time Reference; then, select:
Left — the time reference is set to one major division from the left edge of the
display.
Center — the time reference is set to the center of the display.
Right — the time reference is set to one major division from the right edge of
the display.
Custom Location — lets you place the time reference location at a percent of
the graticule width (where 0% is the left edge and 100% is the right edge).
A small hollow triangle (
) at the top of the display grid marks the position of the
time reference. When delay time is set to zero, the trigger point indicator ()
overlays the time reference indicator.
The time reference position sets the initial position of the trigger event within
acquisition memory and on the display, with delay set to 0.