3 Select Hardware Self Test.
It is recommended you run Hardware Self Test:
After experiencing abnormal operation.
For additional information to better describe an oscilloscope failure.
To verify proper operation after the oscilloscope has been repaired.
Successfully passing Hardware Self Test does not guarantee 100% of the
oscilloscope's functionality. Hardware Self Test is designed to provide an 80%
confidence level that the oscilloscope is operating properly.
To perform front panel self test
To test the front panel keys and knobs, LEDs, display, and touch screen:
1 From the main menu, choose Utilities > Service....
2 In the Service dialog box, select the Diagnostics tab.
3 From the Front Panel Self Test drop-down list, select the test you want to
perform:
Keyboard Test — Tests the front panel keys and knobs.
LED Test — Test the front panel LEDs.
Display Test — Tests the display's ability to output basic colors.
Touch Screen Test — Tests the touch screen.
4 Select Start.
Follow the on-screen instructions.
When the test is complete, status information appears in the Test Summary box.