188 Keysight InfiniiVision HD3-Series Oscilloscopes User's Guide
14 Measurements
5 Specify any other parameters for the selected measurement.
6 Select Add (measurement) to add the measurement selected in the Type list.
Markers are turned on to show the portion of the waveform being measured for
the most recently added measurement (bottom-most on the display). To view
the markers for a previously added measurement (but not the last one), select
the measurement in the Results area.
By default, measurement statistics are displayed. See "Measurement
Statistics"
on page 209.
7 To edit a measurement, select it in the Active Measurements list; then select Edit
(measurement).
8 To show or hide the Measurements window in the Results area, select or clear
On.
9 To delete a measurement, select it in the Active Measurements list; then select
Delete (measurement).
The measurement is removed from the Active Measurements list and from the
Results area.
10 To delete all measurements, select Delete All.
All measurements are removed from the Active Measurements list and from the
Results area.
To edit measurements
When added measurements have parameters that can be edited (like the Delay,
Phase, Occupied Bandwidth, ACPR, or THD measurements), you can edit the
parameters of those measurements.
1 From the menu, choose Measure > Measurements....
2 In the Measurements dialog box, select the Add, Edit & Delete tab.
3 Select the measurement in the Active Measurements list; then select Edit
(measurement).
Measurements Summary
The automatic measurements provided by the oscilloscope are listed in the
following table. All measurements are available for analog channel waveforms. All
measurements except Counter are available for reference waveforms and math
waveforms other than FFT. A limited set of measurements is available for math FFT
waveforms and for digital channel waveforms (as described in the following table).