Triggers 10
Keysight InfiniiVision HD3-Series Oscilloscopes User's Guide 141
Use the > (Gre ater than) field to enter the lower time range value.
Use the < (Less than) field to enter the upper time range value.
The time values can be from 10 ns to 10 s, with minimum difference of 5 ns
between upper and lower settings.
Pattern Trigger
The Pattern trigger identifies a trigger condition by looking for a specified pattern.
This pattern is a logical AND combination of the channels. Each channel can have
a value of 0 (low), 1 (high), or don't care (X). A rising or falling edge can be
specified for one channel included in the pattern.
1 Open the Trigger dialog box by selecting the trigger badge or by choosing
Trigger > Setup... from the main menu.
2 In the Trigger Type drop-down list, select Pattern.
3 From the Qualifier drop-down list, select from the pattern duration qualifier
options:
• Entered — when the pattern is entered.
• < (Less Than) — when the pattern is present for less than a time value.