Performing Service Tasks
The Service dialog box (choose Utilities > Service... from the main menu) lets you
perform service-related tasks:
"To display the user calibration status" on page 287
"To perform user calibration" on page 288
"To perform hardware self test" on page 288
"To perform front panel self test" on page 289
"To display oscilloscope information" on page 290
For other information related to oscilloscope maintenance and service, see:
"To clean the oscilloscope" on page 290
"To check warranty and extended services status" on page 290
"To contact Keysight" on page 290
"To return the instrument" on page 291
To display the user calibration status
To display the summary results of the previous User Calibration and the status of
probe calibrations for probes that can be calibrated:
1 From the main menu, choose Utilities > Service....
2 In the Service dialog box, select the Calibration tab.
Note that passive probes do not need to be calibrated, but InfiniiMax probes can
be calibrated. For more information about calibrating probes see "To calibrate a
probe"
on page 62.