339
Keysight InfiniiVision HD3-Series Oscilloscopes
User's Guide
27 LIN Triggering and Protocol
Decode
LIN Protocol Decode Setup / 339
Loading and Displaying LIN Symbolic Data / 342
LIN Triggering / 342
Interpreting LIN Decode / 344
Interpreting LIN Lister Data / 345
Searching for LIN Data in the Lister / 345
LIN triggering and protocol decode option is included with the Automotive Serial
Analysis license.
LIN Protocol Decode Setup
To enable LIN protocol decode:
1 Press [Protocol Decode].
Or:
a From the main menu, choose Analyze > Protocol Decode....
b In the Protocol Decode dialog box, select Enable.
c Select either the P1 or P2 tab (the oscilloscope can decode two protocols at
the same time), and select On.
2 From the Mode drop-down list, select LIN.
Below the Mode selection are vertical tabs for specifying Signals and Settings.
Signals LIN (Local Interconnect Network) signal setup consists of connecting the
oscilloscope to a serial LIN signal, specifying the signal source, threshold voltage
level, baud rate, sample point, and other LIN signal parameters.
The LIN protocol decode mode Signals subtab has these controls for setting up the
signal: