Histogram 15
Keysight InfiniiVision HD3-Series Oscilloscopes User's Guide 217
Measurement Histogram Set Up
Measurement histograms show the distribution of a measurement's results similar
to ordinary measurement statistics.
To set up a measurement histogram:
1 First, add the measurement whose results you want to display as a histogram.
See "To make automatic measurements" on page 187.
2 Turn on measurement histogram analysis.
You can:
a From the menu, choose Measure > Measurements....
b In the Measurements dialog box, select the Histogram tab.
Or, you can:
a From the menu, choose Analyze > Histogram....
b In the Histogram dialog box, select On.
c From the Type drop-down list, select Measurements.
3 From the Measurement drop-down list, select the measurement to analyze.
4 In the Histogram Size field, specify the number of divisions the histogram bar
graph should use. You can choose whole and half divisions from 1 to 4.
For measurement histograms, information about the scale of the bar graph
appears in the grid. This scale is automatically determined.
You can select Reset Histogram to zero the histogram counters.