144 Keysight InfiniiVision HD3-Series Oscilloscopes User's Guide
10 Triggers
OR Trigger
The OR trigger mode triggers when any one (or more) of the specified edges on
analog or digital channels is found.
1 Open the Trigger dialog box by selecting the trigger badge or by choosing
Trigger > Setup... from the main menu.
2 In the Trigger Type drop-down list, select OR.
3 Use the Set All buttons to set the edges for all channels to (Rising),
(Falling), (Either), or X (Don't Care).
4 To set the edges for individual channels, select the channel; then, select one of
the Set Selected Channel buttons: (Rising), (Falling), (Either), or X (Don't
Care).
If all channels in the OR trigger are set to don't care, the oscilloscope will not
trigger.
5 Select Trigger Levels... to open the Trigger Levels dialog box where you can
adjust the trigger levels for the input channels being used.
Rise/Fall Time Trigger
The Rise/Fall Time trigger looks for a rising or falling transition from one level to
another level in greater than or less than a certain amount of time.