EasyManua.ls Logo

Keysight InfiniiVision HD3 Series - To Change the Logic Threshold for Digital Channels; To Display Digital Channels as a Bus

Keysight InfiniiVision HD3 Series
368 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
110 Keysight InfiniiVision HD3-Series Oscilloscopes User's Guide
7 Digital Channels
Select Descending Order to reorder all channels from higher numbered
channels to lower numbered channels.
Select Ascending Order to reorder all channels from lower numbered channels
to higher numbered channels.
To change the logic threshold for digital channels
1 Open the Digital dialog box by selecting the digital channels badge or by
choosing Sources > Digital Channels....
2 In the Thresholds tab:
Use the D15 - D12, D11 - D8, D7 - D4, and D3 - D0 drop-down lists to select a
logic family preset or select User to define your own threshold.
The threshold you select applies to all channels within the selected D15 - D12,
D11 - D8, D7 - D4, or D3 - D0 group. Each channel group can be set to a different
threshold if desired.
Values greater than the set threshold are high (1) and values less than the set
threshold are low (0).
To display digital channels as a bus
Digital channels may be grouped and displayed as a bus, with each bus value
displayed at the bottom of the display in hex or binary. You can create up to two
buses.
Logic family Threshold Voltage
TTL +1.4 V
CMOS +2.5 V
ECL –1.3 V
User Variable from –8 V to +8 V

Table of Contents