161
 
Keysight InfiniiVision HD3-Series Oscilloscopes 
User's Guide
12 Acquisition Control
Running, Stopping, and Making Single Acquisitions (Run Control) / 161
Overview of Sampling / 162
Selecting the Acquisition Mode / 166
Selecting a Global Bandwidth Limit and Vertical Resolution / 171
Acquiring to Segmented Memory / 171
Digitizer Mode / 174
This chapter shows how to use the oscilloscope's acquisition and run controls.
Running, Stopping, and Making Single Acquisitions (Run Control)
There are two front panel keys for starting and stopping the oscilloscope's 
acquisition system: [Run/Stop] and [Single].
• When the [Run/Stop] key is green, the oscilloscope is running, that is, acquiring 
data when trigger conditions are met.
To stop acquiring data, press [Run/Stop]. When stopped, the last acquired 
waveform is displayed.
• When the [Run/Stop] key is red, data acquisition is stopped.
"Stop" is displayed next to the trigger type in the trigger badge.
To start acquiring data, press [Run/Stop].
• To capture and display a single acquisition (whether the oscilloscope is running 
or stopped), press [Single].
The [Single] run control lets you view single-shot events without subsequent 
waveform data overwriting the display. Use [Single] when you want maximum 
memory depth for pan and zoom.
When you press [Single], the trigger mode is temporarily set to Trig'd (to keep 
the oscilloscope from auto-triggering immediately), the trigger circuitry is 
armed, the [Single] key is illuminated, and the oscilloscope waits until a trigger 
condition occurs before it displays a waveform.