248 Keysight InfiniiVision HD3-Series Oscilloscopes User's Guide
19 Frequency Response Analysis
 
In the Frequency Response dialog box's Chart tab, there are controls for:
• Enabling or disabling the gain or phase plot and table results.
• Adjusting the scroll markers in the plot to see the values measured at particular 
locations.
• Viewing  additional  Chart Controls. The drop-down list lets you select:
• Gain — for controls to adjust the gain plot's vertical scale and offset.
• Phase — for controls to adjust the phase plot's vertical scale and offset.
• Frequency — for controls to adjust the chart's horizontal frequency scale 
(starting and ending frequencies). The Adj Mode drop-down selections let 
you make normal, fine, or decade (coarse) adjustments to the chart's 
starting and ending frequencies.
The Results area table shows: the data point number, the frequency, the waveform 
generator output amplitude, the measured gain, and the measured phase. You can 
scroll the data.
To save the frequency response analysis results:
1 From the main menu, choose File > Save....
2 In the Save dialog box, from the Format drop-down menu, select Frequency 
Response Analysis data (*.csv).
3 Select the File Name field, and enter the name of the file you want to save.
4 Use the controls in the bottom of the dialog box to navigate to the location of 
the file to be saved.
5 Select Save.
To email the frequency response analysis results, see 
"Emailing Setups, Screen 
Images, or Data" on page 267. At the point where you choose the Format of 
information to be emailed, select Frequency Response Analysis data (*.csv).