Waveform Generator 20
Keysight InfiniiVision HD3-Series Oscilloscopes User's Guide 255
a Enter the number of Initial Points.
The new waveform will be a square wave with the number of points you
specify. The points are evenly spaced over the time period.
b Enter the Min Time Between Pts, that is, the minimum time between points.
The resolution is specified during arbitrary waveform creation only. To
change the resolution of an arbitrary waveform, you must clear all arbitrary
waveform points (with a Reset when editing points), and create the waveform
again.
c When you are ready to create the new arbitrary waveform, select Apply &
Edit....
The new waveform is created and the Arbitrary Waveform Editor dialog box
is opened.
• Import waveforms from channel — When selected, to capture another waveform to
the arbitrary waveform:
a From the Source drop-down list, select the analog channel, math, or
reference location whose waveform you wish to capture.
b Select Copy Waveform....
The source waveform is decimated into 8192 (maximum) or fewer arbitrary
waveform points.
If the source waveform frequency and/or voltage exceed the capabilities of the waveform
generator, the arbitrary waveform will be limited to the capabilities of the waveform generator.
For example, a 50 MHz waveform captured as the arbitrary waveform, becomes a 25 MHz
waveform.