CAN Triggering and Protocol Decode 25
Keysight InfiniiVision HD3-Series Oscilloscopes User's Guide 329
• Search for — Selects the decode value to find:
• Frame ID — Finds remote or data frames matching the specified ID.
• Data Frame ID — Finds data frames matching the specified ID.
• Data Frame ID and Data — Finds data frames matching the specified ID and
data.
• Remote Frame ID — Finds remote frames with the specified ID.
• Error Frame — Finds CAN active error frames.
• Acknowledge Error — Finds the acknowledge bit if the polarity is incorrect.
• Form Error — Finds reserved bit errors.
• Stuff Error — Finds 6 consecutive 1s or 6 consecutive 0s, while in a non-error
or non overload frame.
• CRC Field Error — Finds when the calculated CRC does not match the
transmitted CRC.
• All Errors — Finds any form error or active error.
• Overload Frame — Finds CAN overload frames.
When CAN symbolic data is loaded into the oscilloscope (see "Loading and
Displaying CAN Symbolic Data"
on page 323), you can search for:
• Message — A symbolic message.
• Message and Signal — A symbolic message and a signal value.