LIN Triggering and Protocol Decode 27
Keysight InfiniiVision HD3-Series Oscilloscopes User's Guide 341
Sync Break — Select the minimum number of clocks that define a sync break in
your LIN signal.
Settings The LIN protocol decode mode Settings subtab has these controls for decode
settings:
Show Parity — Choose whether to include the parity bits in the identifier field.
a If you want to mask the upper two parity bits, ensure that the Show Parity
check box is not selected.
b If you want to include the parity bits in the identifier field, ensure that the
Show Parity check box is selected.
Display — When LIN symbolic information is loaded, you can select between Hex
and Symbolic display of the decoded data.
Notes The LIN triggering and serial decode option is license-enabled.
If the decode line does not appear on the display, press the [Protocol Decode] key
to turn it on.
If the oscilloscope is stopped, press the [Run/Stop] key to acquire and decode
data.
If the setup does not produce a stable trigger, the LIN signal may be slow
enough that the oscilloscope is Auto triggering. Press the [Mode] key to set the
trigger mode from Auto to Trig'd.
You can use the horizontal [Zoom] window for easier navigation of the decoded
data.
See Also
"Loading and Displaying LIN Symbolic Data" on page 342
"LIN Triggering" on page 342
"Interpreting LIN Decode" on page 344
"Interpreting LIN Lister Data" on page 345
"Searching for LIN Data in the Lister" on page 345