CAUTION
■ Before starting the machine installation, read and full understand the
instructions specified in “6. INSTALLATION" of this Operation Manual.
• The machine shall be installed by well trained and qualified person only.
■ Before connecting any optional device to the machine, read and full
understand “Handling Manual” of the device.
• The connection shall be performed by well trained and qualified person
only.
■ DO NOT give shock or impact to display and operation panel.
• Take care not to give shock or impact to display and operation panel
during cleaning.
■ Clean the machine thoroughly before use of it.
• If the machine is used after long term of storage, rinse using water for one
hour or longer.
• If the machine is used after long term of storage without removing water,
rinse well until dirty water is not drained.
■ Take care for the machine to be kept clean.
• If the machine is contaminated with blood, dialysate or rinsing fluid,
remove it by wiping without delay.
• Perform regular cleaning and disinfection.
■ There are the cases that display and process do not respond to operation of
some SW’s under certain conditions. DO NOT operate the SW’s forcibly.
■ Use the color coded dialysate concentrate container. Non- color coded one
is not applicable to the machine.
■ Take care to avoid infection during handling of disposables because there is
possibility of contact with hepatitis virus and/or contaminated drug or
chemical.
■ Use water, dialysate concentrates and substitute of acceptable qualities for
safe and effective operation of the machine.
■ Gap between rollers and housing of blood pump can be adjusted within
allowable range. The adjustment may be required according to size of
rolling tube.
CAUTION-5