CAUTION-4
■ Injury or death of patient may be caused if the machine is operated,
maintained or calibrated based on any procedure not specified in this
operation manual or if any accessory device not approved by us is used. In
such case, the injury or death is out of our control. The machine shall be
operated, maintained or calibrated by well trained and qualified specialist
based on the procedure specified in this operation manual.
■ Check actual dialysate concentration by meter.
• Check whether dialysate concentration is correct by the test using
osmometer, conductivity meter or flame photometer.
• Check whether dialysate concentration is correctly displayed.
■ Check no residual disinfectant and acid after rinsing completion.
• After disinfection using sodium hypochlorite or peracetic acid; or after
acid rinsing, check whether residue of the disinfectant or acid is
contained in fluid supply and drain lines by the test using test paper or
reagent.
■ DO NOT apply hot rinse without hot rinse option. It might cause breakdown
of the machine.