WARNING
■ DO NOT use any electric wave generating electronic devices in room or
building where the machine is operated.
• DO NOT use electric wave generating electronic devices such as portable
telephone, transceiver and radio controlled toy in hospital to avoid
malfunction of the machine.
• Manage and instruct for the devices not to be brought in and not to be
used in hospital to prevent trouble occurrence.
DO NOT change parameter setting during dialysis excluding the case of
emergency.
• Change parameter setting before connecting the machine to patient.
• Set parameter as intended after full understanding its function.
■ Provide drain port at floor.
• Keep power cord away from floor to avoid electric shock if short-circuit is
caused by water leak.
• Provide drain port at floor near the machine.
Read the manual before starting check and maintenance of the machine.
• Read and full understand the manual before starting check and
maintenance.
• Check and maintain the machine as prescribed in the manual.
• Entrust check and maintenance not prescribed in the manual to
authorized specialist.
■ DO NOT touch any SW unnecessary for the machine operation during
dialysis.
■ DO NOT place the machine at an angle of ≧5°against vertical line under
loading any object on any container support or hunger to prevent falling
down of the machine.
■ DO NOT load any object on any container support or hunger during
transport of the machine.
■ DO NOT place the machine at an angle of ≧10°against vertical line under
unloaded condition to prevent falling down of the machine.
■ DO NOT connect any devices other than approved ones to the machine.
Otherwise, the total leakage current may exceed the acceptable limit.