10-10
Alarm
No.
Alarm name
Detection
condition
Detection
process
Audible alarm
Fluid line
function
BP
SP
stop
CLV
close
Alarm
release
Resetting and note
E40 VPa CLOSED CIRCUIT
LEAK ERROR
Insufficient positive
pressure in A chamber
No arrival at set pressure
within set time
Rinse(water
rinse)
○(7times) ― ― ― ○
Return to normal by pressing [RESET] SW
Possible alarm release in maintenance
mode 2
E41 VPa CLOSED CIRCUIT
LEAK ERROR
Fluctuated positive
pressure in A chamber
Pressure fluctuation out of
set range during set time
Rinse(water
rinse)
○(7times) ― ― ― ○
Return to normal by pressing [RESET] SW
Possible alarm release in maintenance
mode 2
E42 VPb CLOSED CIRCUIT
LEAK ERROR
Insufficient positive
pressure in B chamber
No arrival at set pressure
within set time
Rinse(water
rinse)
○(7times) ― ― ― ○
Return to normal by pressing [RESET] SW
Possible alarm release in maintenance
mode 2
E43 VPb CLOSED CIRCUIT
LEAK ERROR
Fluctuated positive
pressure in B chamber
Pressure fluctuation out of
set range during set time
Rinse(water
rinse)
○(7times) ― ― ― ○
Return to normal by pressing [RESET] SW
Possible alarm release in maintenance
mode 2
E44 AIR ELIMINATION TANK
ERROR
Functioning of "FSW" of
air elimination tank for ≧
15 sec in dialysis process
or for ≧ 30 sec in any
process other than
dialysis
•Rinse
•Preparation
•Preparation end
•Dialysis
•Recovery
•○(7 times)
•○(Continuous)
•Same as above
•Same as above
•Same as above
•―
•Same as above
•Same as above
•Fluid line stop
•―
― ― ○
Return to normal by pressing [RESET] SW
E45 BLOWN-OUT FUSE
DETECTION
Blown out of pump fuse
•Rinse wait
•Rinse
•Rinse end
•Air purge
•Dialysis wait
•Preparation
•Preparation end
•Dialysis
•Recovery
•○(7times)
•Same as above
•Same as above
•Same as above
•○(Continuous)
•Same as above
•Same as above
•Same as above
•Same as above
Fluid line stop
― ― ―
Return to normal by pressing [RESET] SW
E46 WATCH-DOG ERROR System resetting after
system error detection by
watch dog
•Rinse wait
•Rinse
•Rinse end
•Air purge
•Dialysis wait
•Preparation
•Preparation end
•Dialysis
•Recovery
•○(7times)
•Same as above
•Same as above
•Same as above
•○(Continuous)
•Same as above
•Same as above
•Same as above
•Same as above
― ― ― ―
Return to normal by pressing [RESET] SW
RAM initialization after error detection