10-9
Alarm
No.
Alarm name
Detection
condition
Fluid line
function
BP
SP
stop
CLV
close
Alarm
release
Resetting and note
Detection
process
Audible alarm
E32 CLV POSITION ERROR Abnormal clamp position
•Rinse wait
•Rinse
•Rinse end
•Air purge
•Dialysis wait
•Preparation
•Preparation end
•Dialysis
•Recovery
•○(7times)
•Same as above
•Same as above
•Same as above
•○(Continuous)
•Same as above
•Same as above
•Same as above
•Same as above
― ○
SW-OF
F
― ―
Return to normal by pressing [RESET] SW
Return to normal by BP/SPSW-ON
E33
V
ISCOUS PUMP DRIVING
ERROR(VPa)
Error of viscous pump a
•Rinse
•Preparation
•Preparation end
•Dialysis
•Recovery
•○(7times)
•○(Continuous)
•Same as above
•Same as above
•Same as above
Fluid line stop
― ― ○
Return to normal by pressing [RESET] SW
E34
V
ISCOUS PUMP DRIVING
ERROR(VPb)
Error of viscous pump b
•Rinse
•Preparation
•Preparation end
•Dialysis
•Recovery
•○(7times)
•○(Continuous)
•Same as above
•Same as above
•Same as above
Fluid line stop
― ― ○
[Return to normal by pressing [RESET] SW
E35 VP DRAWING VOLUME
IS AT MAX LIM.
Viscous pump operation
out of allowable range
•Rinse
•Preparation
•Preparation end
•Dialysis
•Recovery
•○(7times)
•○(Continuous)
•Same as above
•Same as above
•Same as above
Fluid line stop
― ― ○
Return to normal by pressing [RESET] SW
E36 VPa CLOSED CIRCUIT
LEAK ERROR
Insufficient negative
pressure in A chamber
No arrival at set pressure
within set time
Rinse(water
rinse)
○(7times) ― ― ― ○
Return to normal by pressing [RESET] SW
Possible alarm release in maintenance
mode 2
E37 VPa CLOSED CIRCUIT
LEAK
ERROR
Fluctuated negative
pressure in A chamber
Pressure fluctuation out of
set range during set time
Rinse(water
rinse)
○(7times) ― ― ― ○
Return to normal by pressing [RESET] SW
Possible alarm release in maintenance
mode 2
E38 VPb CLOSED CIRCUIT
LEAK
ERROR
Insufficient negative
pressure in B chamber
No arrival at set pressure
within set time
Rinse(water
rinse)
○(7times) ― ― ― ○
Return to normal by pressing [RESET] SW
Possible alarm release in maintenance
mode 2
E39 VPb CLOSED CIRCUIT
LEAK
ERROR
Fluctuated negative
pressure in B chamber
Pressure fluctuation out of
set range during set time
Rinse(water
rinse)
○(7times) ― ― ― ○
Return to normal by pressing [RESET] SW
Possible alarm release in maintenance
mode 2