EasyManua.ls Logo

Nipro SURDIAL - Page 83

Default Icon
198 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10-9
Alarm
No.
Alarm name
Detection
condition
Fluid line
function
BP
SP
stop
CLV
close
Alarm
release
Resetting and note
Detection
process
Audible alarm
E32 CLV POSITION ERROR Abnormal clamp position
Rinse wait
Rinse
Rinse end
Air purge
Dialysis wait
Preparation
Preparation end
Dialysis
Recovery
(times)
Same as above
Same as above
Same as above
(Continuous)
Same as above
Same as above
Same as above
Same as above
SW-OF
F
Return to normal by pressing [RESET] SW
Return to normal by BP/SPSW-ON
E33
V
ISCOUS PUMP DRIVING
ERROR(VPa)
Error of viscous pump a
Rinse
Preparation
Preparation end
Dialysis
Recovery
(times)
(Continuous)
Same as above
Same as above
Same as above
Fluid line stop
Return to normal by pressing [RESET] SW
E34
V
ISCOUS PUMP DRIVING
ERROR(VPb)
Error of viscous pump b
Rinse
Preparation
Preparation end
Dialysis
Recovery
(times)
(Continuous)
Same as above
Same as above
Same as above
Fluid line stop
[Return to normal by pressing [RESET] SW
E35 VP DRAWING VOLUME
IS AT MAX LIM.
Viscous pump operation
out of allowable range
Rinse
Preparation
Preparation end
Dialysis
Recovery
(times)
(Continuous)
Same as above
Same as above
Same as above
Fluid line stop
Return to normal by pressing [RESET] SW
E36 VPa CLOSED CIRCUIT
LEAK ERROR
Insufficient negative
pressure in A chamber
No arrival at set pressure
within set time
Rinse(water
rinse)
(times)
Return to normal by pressing [RESET] SW
Possible alarm release in maintenance
mode 2
E37 VPa CLOSED CIRCUIT
LEAK
ERROR
Fluctuated negative
pressure in A chamber
Pressure fluctuation out of
set range during set time
Rinse(water
rinse)
(times)
Return to normal by pressing [RESET] SW
Possible alarm release in maintenance
mode 2
E38 VPb CLOSED CIRCUIT
LEAK
ERROR
Insufficient negative
pressure in B chamber
No arrival at set pressure
within set time
Rinse(water
rinse)
(times)
Return to normal by pressing [RESET] SW
Possible alarm release in maintenance
mode 2
E39 VPb CLOSED CIRCUIT
LEAK
ERROR
Fluctuated negative
pressure in B chamber
Pressure fluctuation out of
set range during set time
Rinse(water
rinse)
(times)
Return to normal by pressing [RESET] SW
Possible alarm release in maintenance
mode 2

Table of Contents

Other manuals for Nipro SURDIAL

Related product manuals