10-8
Alarm
No.
Alarm name
Detection
condition
Audible alarm
Fluid line
function
BP
SP
stop
CLV
close
Alarm
release
Resetting and note
Detection
process
E25 HEATER OVER HEAT
ALARM
Functioning of "LS1"
placed at heater outside
wall
Preparation
Preparation end
Dialysis
Recovery
(Continuous)
Heater OFF
Return to normal by pressing [RESET] SW
E26 SHORT WATER SUPPLY Detection of no water
supply pressure by
pressure SW "PS"
Rinse
Preparation
Preparation end
Dialysis
Recovery
(times)
(Continuous)
Same as above
Same as above
Same as above
Fluid line stop
[Return to normal by pressing [RESET] SW
E27 SHORT DIALYSATE
FLOW
No functioning of "FS" for
85 sec after chamber
change-over
Preparation
Preparation end
Dialysis
Recovery
(Continuous)
[Return to normal by pressing [RESET] SW
E28 DIALYSATE FLOW STOP
Functioning of “FS”
within 20 sec after
chamber change-over
Functioning of “FS” for
170 sec during dialysis
process other than SQHD
Rinse(water
rinse)
Preparation
Preparation end
Dialysis
Recovery
(times)
(Continuous)
Same as above
Same as above
Same as above
Fluid line stop
[Return to normal by pressing [RESET] SW
E29 SN SWITCH TIME ALARM No arrival at changeover
pressure after set SN
alarm time
Dialysis
(Continuous)
UF stop
[Return to normal by pressing [RESET] SW
Error detection only when SN SW is ON
E30 BLOOD PUMP ERROR
Overload
Abnormal rotating
direction
Abnormal rotating speed
Rinse wait
Rinse
Rinse end
Air purge
Dialysis wait
Preparation
Preparation end
Dialysis
Recovery
(times)
Same as above
Same as above
Same as above
(Continuous)
Same as above
Same as above
Same as above
Same as above
SW-OF
F
Return to normal by pressing [RESET] SW
Return to normal by BP/SPSW-ON
E31 BLOOD PUMP STOP
ERROR
Opening of blood pump
over
OFF of blood pump
[ON/OFF] SW
0 setting of blood pump
flow rate setting volume
Dialysis
(Continuous)
UF stop SPstop
Return to normal by pressing [RESET] SW
Auto reset by release from detection
condition