10-7
Alarm
No.
Alarm name
Detection
condition
Audible alarm
Fluid line
function
BP
SP
stop
CLV
close
Alarm
release
Resetting and note
Detection
process
E17 ARTERIAL PRESSURE
MAX ALARM
Arterial pressure of set
upper alarm limit
Dialysis wait
Preparation
Preparation end
Dialysis
Recovery
(Continuous)
UF stop
Return to normal by pressing [RESET] SW
Becoming error free of set upper and lower
DP alarm limits by alarm actuation
E18 ARTERIAL PRESSURE
MIN ALARM
Arterial pressure of set
lower alarm limit
Dialysis wait
Preparation
Preparation end
Dialysis
Recovery
(Continuous)
UF stop
Return to normal by pressing [RESET] SW
Becoming error free of set upper and lower
DP alarm limits by alarm actuation
E19 TMP ALARM
[VP-DP] of set alarm
limit
Dialysis
Recovery
(Continuous)
UF stop
Return to normal by pressing [RESET] SW
E20 DIALYSATE
CONDUCTIVITY
MAX ALARM
Dialysate conductivity of
set upper alarm limit
Dialysis
(Continuous)
Auto bypass
Return to normal by pressing [RESET] SW
Auto reset after release from detection
condition for 10 sec
E21 DIALYSATE
CONDUCTIVITY MIN
ALARM
Dialysate conductivity of
set lower alarm limit
Dialysis
(Continuous)
Auto bypass
Return to normal by pressing [RESET] SW
Auto reset after release from detection
condition for 10 sec
E22 BLOOD LEAK SENSOR
ERROR
Detection of absolute
blood leak REAL value of
201 in preparation or
preparation end process
Detection of absolute
blood leak REAL value of
201 in dialysis or
recovery process
Signaling of rinse alarm
Preparation
Preparation end
Dialysis
Recovery
(Continuous)
Fluid line stop
Return to normal by pressing [RESET] SW
E23 BLOOD LEAK ALARM
Blood leak value of set
alarm limit
Dialysis
(Continuous)
UF stop
Auto bypass
Return to normal by pressing [RESET] SW
E24 AIR BUBBLES ARE
DETECTED
Bubble detection during
ON of BUBBLE [ON/OFF]
SW
Rinse wait
Rinse
Rinse end
Air purge
Dialysis wait
Preparation
Preparation end
Dialysis
Recovery
(7 times)
Same as above
Same as above
Same as above
(Continuous)
Same as above
Same as above
Same as above
Same as above
UF stop
SW-OF
F
Return of UF stop/CLV to normal by
pressing [RESET] SW
Return of BP/SP to normal by SW-ON
Error detection only when BUBBLE
[ON/OFF] SW is ON