10-7
Alarm
No.
Alarm name
Detection
condition
Audible alarm
Fluid line
function
BP
SP
stop
CLV
close
Alarm
release
Resetting and note
Detection
process
E17 ARTERIAL PRESSURE
MAX ALARM
Arterial pressure of ≧set
upper alarm limit
•Dialysis wait
•Preparation
•Preparation end
•Dialysis
•Recovery
○(Continuous)
UF stop
○ ― ○
Return to normal by pressing [RESET] SW
Becoming error free of set upper and lower
DP alarm limits by alarm actuation
E18 ARTERIAL PRESSURE
MIN ALARM
Arterial pressure of ≦set
lower alarm limit
•Dialysis wait
•Preparation
•Preparation end
•Dialysis
•Recovery
○(Continuous)
UF stop
○ ― ○
Return to normal by pressing [RESET] SW
Becoming error free of set upper and lower
DP alarm limits by alarm actuation
E19 TMP ALARM
[VP-DP] of ≧ set alarm
limit
•Dialysis
•Recovery
○(Continuous)
UF stop
○ ○ ○
Return to normal by pressing [RESET] SW
E20 DIALYSATE
CONDUCTIVITY
MAX ALARM
Dialysate conductivity of
≧set upper alarm limit
Dialysis
○(Continuous)
Auto bypass
― ― ―
Return to normal by pressing [RESET] SW
Auto reset after release from detection
condition for 10 sec
E21 DIALYSATE
CONDUCTIVITY MIN
ALARM
Dialysate conductivity of
≦set lower alarm limit
Dialysis
○(Continuous)
Auto bypass
― ― ―
Return to normal by pressing [RESET] SW
Auto reset after release from detection
condition for 10 sec
E22 BLOOD LEAK SENSOR
ERROR
•Detection of absolute
blood leak REAL value of
≧ 201 in preparation or
preparation end process
•Detection of absolute
blood leak REAL value of
≦ 201 in dialysis or
recovery process
•Signaling of rinse alarm
•Preparation
•Preparation end
•Dialysis
•Recovery
○(Continuous)
Fluid line stop
― ― ○
Return to normal by pressing [RESET] SW
E23 BLOOD LEAK ALARM
Blood leak value of ≧set
alarm limit
Dialysis
○(Continuous)
UF stop
Auto bypass
○ ― ○
Return to normal by pressing [RESET] SW
E24 AIR BUBBLES ARE
DETECTED
Bubble detection during
ON of BUBBLE [ON/OFF]
SW
•Rinse wait
•Rinse
•Rinse end
•Air purge
•Dialysis wait
•Preparation
•Preparation end
•Dialysis
•Recovery
•○(7 times)
•Same as above
•Same as above
•Same as above
•○(Continuous)
•Same as above
•Same as above
•Same as above
•Same as above
UF stop
○
SW-OF
F
○ ○
Return of UF stop/CLV to normal by
pressing [RESET] SW
Return of BP/SP to normal by SW-ON
Error detection only when BUBBLE
[ON/OFF] SW is ON