10-6
Alarm
No.
Alarm name
Detection
condition
Fluid line
function
BP
SP
stop
CLV
close
Alarm
release
Resetting and note
Detection
process
Audible alarm
E12
V
ENOUS PRESSURE MIN
LIMIT
Venous pressure of
-400mmHg
Rinse wait
Rinse
Rinse end
Air purge
Dialysis wait
Preparation
Preparation end
Dialysis
Recovery
(7 times)
Same as above
Same as above
Same as above
(Continuous)
Same as above
Same as above
Same as above
Same as above
UF stop
Return to normal by pressing [RESET] SW
E13 VENOUS PRESSURE
MAX ALARM
Venous pressure of set
upper alarm limit
Dialysis wait
Preparation
Preparation end
Dialysis
Recovery
(Continuous)
UF stop
Return to normal by pressing [RESET] SW
Becoming error free of set upper and lower
DP alarm limits by alarm actuation
E14
V
ENOUS PRESSURE MIN
ALARM
Venous pressure of set
lower alarm limit
Dialysis wait
Preparation
Preparation end
Dialysis
Recovery
(Continuous)
UF stop
Return to normal by pressing [RESET] SW
Becoming error free of set upper and lower
VP alarm limits by alarm actuation
E15 ARTERIAL PRESSURE
MAX LIMIT
Arterial pressure of
+400mmHg
Rinse wait
Rinse
Rinse end
Air purge
Dialysis wait
Preparation
Preparation end
Dialysis
Recovery
(times)
Same as above
Same as above
Same as above
(Continuous)
Same as above
Same as above
Same as above
Same as above
UF stop
Return to normal by pressing [RESET] SW
E16 ARTERIAL PRESSURE
MIN LIMIT
Arterial pressure of
-400mmHg
Rinse wait
Rinse
Rinse 終了
Air purge
Dialysis wait
Preparation
Preparation end
Dialysis
Recovery
(times)
Same as above
Same as above
Same as above
(Continuous)
Same as above
Same as above
Same as above
Same as above
UF stop
[Return to normal by pressing [RESET] SW