10-12
Alarm
No.
Alarm name
Detection
condition
Fluid line
function
BP
SP
stop
CLV
close
Alarm
release
Resetting and note
Detection
process
Audible alarm
E51 EEPROM INITIALIZE EEPROM initialization
•Rinse wait
•Rinse
•Rinse end
•Air purge
•Dialysis wait
•Preparation
•Preparation end
•Dialysis
•Recovery
•○(7times)
•Same as above
•Same as above
•Same as above
•○(Continuous)
•Same as above
•Same as above
•Same as above
•Same as above
― ― ― ―
Return to normal by pressing [RESET] SW
Error detection only when power SW is ON
or power failure is corrected
Auto shift to dialysis wait process after
error detection
E52 SWITCH ERROR ON of any SW on
operation panel for 30 sec
continuously at power SW
ON
•Rinse wait
•Rinse
•Rinse end
•Air purge
•Dialysis wait
•Preparation
•Preparation end
•Dialysis
•Recovery
•○(7times)
•Same as above
•Same as above
•Same as above
•○(Continuous)
•Same as above
•Same as above
•Same as above
•Same as above
― ― ― ―
Return to normal by pressing [RESET] SW
E53 I/O PORT ERROR Error of I/O port on board
•Rinse wait
•Rinse
•Rinse end
•Air purge
•Dialysis wait
•Preparation
•Preparation end
•Dialysis
•Recovery
•○(7times)
•Same as above
•Same as above
•Same as above
•○(Continuous)
•Same as above
•Same as above
•Same as above
•Same as above
Fluid line stop
○ ― ―
Return to normal by power SW ON
Disabled all functions of the machine and
components on operation panel after error
detection
E54 CONNECTION ERROR
SBY0151-SBY0156
Disconnection of harness
used to connect main
board to operation board
•Rinse wait
•Rinse
•Rinse end
•Air purge
•Dialysis wait
•Preparation
•Preparation end
•Dialysis
•Recovery
•○(7times)
•Same as above
•Same as above
•Same as above
•○(Continuous)
•Same as above
•Same as above
•Same as above
•Same as above
Fluid line stop
○ ― ―
Return to normal by power SW ON
Disabled all functions of the machine and
components on operation panel after error
detection