10-13
Alarm
No.
Alarm name
Detection
condition
Fluid line
function
BP
SP
stop
CLV
close
Alarm
release
Resetting and note
Detection
process
Audible alarm
E55 BP BATTERY ERROR BP backup battery voltage
of 8.5V
Rinse wait
Rinse
Rinse end
Air purge
Dialysis wait
Preparation
Preparation end
Dialysis
Recovery
(times)
Same as above
Same as above
Same as above
(Continuous)
Same as above
Same as above
Same as above
Same as above
Return to normal by pressing [RESET] SW
Possible alarm release in maintenance
mode 7
E56 COUPLER SW ON
ALARM (DIALYSIS)
"LS2" ON Dialysis
(Continuous)
Fluid line stop
Return to normal by pressing [RESET] SW
E57 COUPLER SW ON
ALARM (PREP.)
Pressing [UF DIALYSE]
SW during "LS2" is ON
Preparation end
Recovery
(3 times)
Return to normal by pressing [RESET] SW
Auto reset by release from detection
condition
E58 COUPLER SW OFF
ALARM
"LS2" OFF Rinse
(3 times)
Fluid line stop
Return to normal by pressing [RESET] SW
Auto reset by release from detection
condition
E59 UF MAX ALARM BP flow rate/UFR relation
of set upper UFR alarm
limit
Dialysis
(Continuous)
UF stop
Return to normal by pressing [RESET] SW
Auto reset by release from detection
condition
E60 SYRINGE INFUSION
COMPLETION
Syringe pump SW ON; and
detection of the pump
over-load signal "LS3"
Rinse wait
Rinse
Rinse end
Air purge
Dialysis wait
Preparation
Preparation end
Dialysis
Recovery
(times)
Same as above
Same as above
Same as above
(Continuous)
Same as above
Same as above
Same as above
Same as above
SP stop
Return to normal by pressing [RESET] SW
Auto reset by release from detection
condition
Error detection when SP SW is ON
E61 COMPLETION OF
SYRINGE.P INFUSION
Arrival at preset syringe
operation stop time
Dialysis
(3 times)
SP stop
Return to normal by pressing [RESET] SW
Auto reset by release from detection
condition
E62 PERFORMED PRIMING Priming completion with
preset solution volume
Priming
(3 times)
BP stop
SW-OF
F
Return to normal by pressing [RESET] SW
Auto reset by release from detection
condition
E63 PREFORMED TARGET
UF.
Achievement of target UF
volume
Dialysis
(Continuous)
UF stop
Return to normal by pressing [RESET] SW
Auto reset by release from detection
condition
Becoming error free of set upper and lower
VP/DP alarm limits by alarm actuation