10-14
Alarm
No.
Alarm name
Detection
condition
Detection
process
Audible alarm
Fluid line
function
BP
SP
stop
CLV
close
Alarm
release
Resetting and note
E64 NO WATER RINSING Shift to dialysis wait
process without executing
water rinse for specified
time after
disinfection/rinse, acid
rinse, hot water rinse or
hot chemical disinfectant
rinse.
Dialysis wait
○(3 times) ― ― ― ―
Return to normal by pressing [RESET] SW
Return to normal by water rinsing for
specified time
E65 PREPARE DIALYSATE
CONC.
Detection of no pressure
rise in A or B concentrate
line by "PS" and detection
of no connection of the
line to the machine
Preparation
○(Continuous)
Fluid line stop
― ― ―
Return to normal by pressing [RESET] SW
E66 CONC CONNECTOR TO
RINSING PORT.
Detection of pressure rise
in A or B concentrate line
by "PS" and detection of
no connection of the line
to rinse port
Rinse
Preparation
○ (3 times)
○ (continuour)
Fluid line stop
― ― ―
Return to normal by pressing [RESET] SW
E67 RESET SW No press of [RESET] SW
for set time after alarm
actuation
•Dialysis wait
•Preparation
•Preparation end
•Dialysis
•Recovery
○(Continuous) ― ― ― ○
Return to normal by pressing [RESET] SW
Possible alarm release in maintenance
mode 2
E68 UF DIALYSE SW No press of [UF DIALYSE]
SW for set time
Preparation end
○(3 times) ― ― ― ○
Return to normal by pressing [RESET] SW
Possible alarm release in maintenance
mode 2
E69 BYPASS SW No release from manual
ON of [BYPASS] SW
within set time
Dialysis
○(3 times) ― ― ― ○
Return to normal by pressing [RESET] SW
Possible alarm release in maintenance
mode 2