APPLIED OPERATIONS/Image Acquisition
OPERATION INSTRUCTIONS
2-18
Page
One Point!
After the dyeing method has been set with the [Dyes] sub-panel, the ND filters
can be set using the [Ch] group box in the upper part of the [Acquire] panel.
1. In the upper part of the [Acquire] sub-panel, open the [Ch] group box for the
ND to be changed.
2. Click the <More> button.
The field as shown below is displayed below the [Ch] group box.
3. Vary the ND value using the Laser LED slider.
Each click of the laser ND <+> or <-> varies the laser ND value by 1%.
Each click of the laser ND slider varies the laser ND value by 5%.
Clicking this field allows the ND value
to be changed on a large scale.
The ND values which are usually
used are displayed in green.
Clicking this button allows fine
adjustment of the ND value.
Display of the ND value set by
clicking the <+> and <-> buttons o
the field. The set ND value can also
be changed by entering its value
directly from the keyboard.
Display of the optimum laser and
ND value which are set
automatically according to the
selected dyeing method.